నైరూప్య

Possibilities Of Using A New Graphite-Castor Oil Polyurethane Composite Electrode As An Amperometric Flow Detector

????????der Tadeu Gomes Cavalheiro, Priscila Cervini, Luiz Ant???????´nio Ramos


The performance of a new graphite castor oil polyurethane composite electrode is evaluated as an amperometric flow injection detector using hydroquinone (HQ) as a probe. HQ was determined in photographic developers with a minimum detectable net concentration of 100 μmol L-1, and 81 determinations h-1, after optimization of parameters such as the detection potential, sample loop volume, and carrier solution flow rate. Interference of SO3 2- present in the developers has been observed and could be eliminated by appropriate sample preparation. Results obtained at the composite electrode agreed with those from HPLC procedure within 95% confidence level, showing that the new composite can be used as an amperometric detector for flow analysis.


ఇండెక్స్ చేయబడింది

  • CASS
  • గూగుల్ స్కాలర్
  • J గేట్ తెరవండి
  • చైనా నేషనల్ నాలెడ్జ్ ఇన్‌ఫ్రాస్ట్రక్చర్ (CNKI)
  • CiteFactor
  • కాస్మోస్ IF
  • ఎలక్ట్రానిక్ జర్నల్స్ లైబ్రరీ
  • రీసెర్చ్ జర్నల్ ఇండెక్సింగ్ డైరెక్టరీ (DRJI)
  • రహస్య శోధన ఇంజిన్ ల్యాబ్‌లు
  • ICMJE

మరిన్ని చూడండి

జర్నల్ హెచ్-ఇండెక్స్

Flyer