నైరూప్య

Calorimetric determination of the neutralization of tham-HCl from 00C to 500C using a new isothermal calorimeter connected to an automated temperature transducer

J.C.Moreno-Piraj???????¡n, L.Giraldo-Guti???????©rrez


At our laboratory, we developed a new cell type for the batch isothermal calorimetry of dissolution enthalpies to determine small amounts of easily soluble or slightly soluble solids. An innovative mixing system aimed at minimizing errors related to the common brittle point breakage effect has been coupled with a pneumatic valve, thus allowing it to work as a cell within an environment involving automatic information management done with software and electronics. The calorimetric cell has a capacity of 20 mL and the sample valve can handle solid samples between 5.0 and 30.0 mg. The experimental conditions suggested in specialized literature were maintained. Values of -H° of 33.41340.0005, 32.1248 0.0012, 30.7357 0.0014, 30.2754  0.0017, 30.2691  0.0007, 29.33950. 0023, 28.7232  0.0032 and 26.0705  0.0028Kjoul./mole at 0, 10, 10, 15, 20, 25, 30, 40, and 50°C, respectively, are reported. The results of the study are then compared with existing studies in calorimetrics, and confirmthe results of the standard THAMHCl calorimetric system.


ఇండెక్స్ చేయబడింది

  • CASS
  • గూగుల్ స్కాలర్
  • J గేట్ తెరవండి
  • చైనా నేషనల్ నాలెడ్జ్ ఇన్‌ఫ్రాస్ట్రక్చర్ (CNKI)
  • CiteFactor
  • కాస్మోస్ IF
  • ఎలక్ట్రానిక్ జర్నల్స్ లైబ్రరీ
  • రీసెర్చ్ జర్నల్ ఇండెక్సింగ్ డైరెక్టరీ (DRJI)
  • రహస్య శోధన ఇంజిన్ ల్యాబ్‌లు
  • ICMJE

మరిన్ని చూడండి

జర్నల్ హెచ్-ఇండెక్స్

Flyer